PERBEDAAN KUALITAS CITRA DIGITAL RADIOGRAFI (DR) WRIST JOINT PROYEKSI PA PADA BUCKY TABLE DAN BUCKY STAND
DOI:
https://doi.org/10.31004/joecy.v6i2.13938Keywords:
Digital Radiography, Wrist Joint, Noise, SNR, CNRAbstract
In Digital Radiography (DR) examinations of the wrist joint using the PA projection, image quality can be compromised by the presence of noise, which reduces the clarity of anatomical details. Research comparing the image quality of PA-projection wrist joint DR examinations performed using a bucky table versus a bucky stand remains limited. This study aimed to evaluate the image quality of PA-projection wrist joint radiographs obtained using bucky tables and bucky stands—based on Signal-to-Noise Ratio (SNR) and Contrast-to-Noise Ratio (CNR) values—and to identify the factors contributing to noise. This mixed-methods study was conducted at Dr. Gunawan Mangunkusumo Regional General Hospital (RSUD) between March and April 2026. Data were collected through observation, interviews, documentation review, and Region of Interest (ROI) measurements using the RadiAnt DICOM Viewer application. Observational, interview, and documentation data were analyzed descriptively, while ROI data were analyzed based on noise levels, SNR, and CNR to objectively assess image quality. The study found that using a bucky stand resulted in a lower noise value (362.81) compared to the bucky table (956.42). SNR and CNR values were also higher for the bucky stand (113.23 and 62.42, respectively) than for the bucky table (60.35 and 26.45). Factors contributing to noise in PA-projection wrist joint DR images taken on the bucky table included the use of lower exposure factors (42.1 kV and 2.25 mAs), the condition of the radiographic equipment, and the digital image processing workflow. The study concludes that using a bucky stand for PA-projection wrist joint DR examinations yields superior image quality compared to using a bucky table, as evidenced by the noise, SNR, and CNR values. The use of a bucky stand may be considered to achieve optimal image quality, and radiographers need to pay attention to exposure factor settings to minimize noise.
References
Abdi, A. J., Precht, H., Outzen, C. B., & Jensen, J. (2025). Quantitative Evaluation of Image Quality and Radiation Dose Using Novel Intelligent Noise Reduction ( INR ) Software in Chest Radiography : A Phantom Study. 1–16.
Anam, C., Triadyaksa, P., Naufal, A., Muhlisin, Z., Setiawati, E., & Budi, W. S. (2022). Impact of ROI Size on the Accuracy of Noise Measurement in CT on Computational and ACR Phantoms. February, 359–368. https://doi.org/10.31661/jbpe.v0i0.2202-1457.Keywords
Angella, S. (2025). Optimalisasi Kualitas Citra Radiograf Menggunakandan Digital Radiography ( DR ) Pada Pemeriksaan Anatomi Thorax Proyeksi Antero - Posterior ( AP ). 5, 5885–5895.
Astria, R., Lasiyah, N., Mulyadi, R., Al, T., & Ningtias, R. (2024). ANALISIS KUALITAS CITRA RADIOGRAFI CR DENGAN SIGNAL TO NOISE RATIO ( SNR ) DAN CONTRAS TO NOISE RATIO ( CNR ) MENGGUNAKAN MICRODICOM Sinar X yang dimanfaatkan dalam bidang medis semakin berkembang dan meningkat seiring dengan perkembangan teknologi .
Barai, J., Tivaskar, S., Sable, A., & Luharia, A. (2022). D. O. and I. Q. in D. R. (2022). Dose Optimization and Image Quality in Digital Radiography : A Review Article. 13(3), 1076–1080. https://doi.org/10.47750/pnr.2022.13.03.174
Bontrager, K.L., & Lampignano, J. P. (2018). T. of R. P. and R. A. (9th ed. ). E. (n.d.). TEXTBOOK OF POSITIONING AND RELATED.
Emanuel, P., Setiawan, D., Hartoyo, P., Widodo, S., Larasati, N., & Sari, K. (2025). Evaluasi Nilai Signal to Noise Ratio ( SNR ) dengan Penerapan Aturan 10 kVp dan 15 % kVp pada Objek Stepwedge. 28(1), 1–10.
Michail, C., Liaparinos, P., & Kalyvas, N. (2024). Radiation Detectors and Sensors in Medical Imaging.
Mustafa, M. R., Ade, I., Liscyaningsih, N., & Widyasari, D. (2025). ANALISIS PERBANDINGAN KUALITAS CITRA RADIOGRAFI ABDOMEN DENGAN MENGGUNAKAN TEKNOLOGI VIRTUAL GRID. 6, 267–273.
Pelawi, A., Efarina, U., Digital, R., Citra, K., Anatomi, D., & Wavelet, T. (2025). Pengembangan Algoritma Denoising Citra Radiografi Digital untuk Peningkatan Kontras dan Detil Anatomi. 2–4.
Susanto, F., Sedya, H., Aulia, U., Alchamdani, R., Nisa, A., & Fatimah, R. (2018). Penerapan Modifikasi Faktor Eksposi Aturan 10 kV Terhadap Kualitas Citra Radiografi Thorax Modalitas CR vs DR. 1–57.
Tsalafoutas, I. A., Alkhazzam, S., Tsapaki, V., Alnaemi, H., & Kharita, M. H. (2022). Digital radiography image quality evaluation using various phantoms and software. June, 1–15. https://doi.org/10.1002/acm2.13823
Tsarouchi, A. M. I., Portaluri, A., Maas, M., & Mann, R. M. (n.d.). Title : Comparing DWI image quality of deep-learning-reconstructed EPI with RESOLVE in breast lesions at 3 . 0T : a pilot study.
Downloads
Published
How to Cite
Issue
Section
License
Copyright (c) 2026 Febtya Nurzaharah, Ildsa Maulidya N, Anisa Istiqomah

This work is licensed under a Creative Commons Attribution-ShareAlike 4.0 International License.